Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems.

Streichert T, Otto B, Schnabel C, Nordholt G, Haddad M, Maric M, Petersmann A, Jung R, Wagener C

Publication: Clin Chem (2011) • Vol 57 • Page 1390-7PMID: 21836074View on PubMed

Purpose

The purpose of this paper was to determine the effects of transporting specimens via pneumatic tube system (PTS) and PTS speed on 42 clinical chemistry and hematological analytes in blood.

Conclusion

After PTS at 2.5 m/s potassium (K), phosphate, aspartate aminotransferase (ASAT) and lactate dehydrogenase (LDH) increased beyond the allowance of the guidelines of the German Federal Medical Council. The magnitude of the changes decreased with decreasing transport speed. The changes in LDH, ASAT and K were correlated with hemolysis which also increased with increasing transport speed.

Studies (1)

Purpose

The purpose of this study was to determine the effects of transporting specimens via PTS and PTS speed on 42 clinical chemistry and hematological analytes in blood. Specimens were collected into K EDTA, coagulation, or ammonium heparin tubes, serum monovettes, or PICO samplers as appropriate. Transport by PTS and hand-delivery took 5-6.5 min and 7.5 min, respectively and there were no difference in specimen transport, humidity or pressure during transport, but significant differences in 3-axis acceleration occurred.

Summary of Findings

PTS transport at 2.5 m/s significantly affected platelet and monocyte counts, reptilase time, D-dimer fragments, antithrombin, K, phosphorus, magnesium, ASAT, alanine aminotransferase (ALAT), lipase, and LDH levels, but only the increases in K, phosphate ASAT and LDH exceeded the allowance of the guidelines of the German Federal Medical Council. The magnitude of the changes, which were not always directionally consistent, decreased with decreasing transport speed, and, at a speed of 1.5 m/s, only the increases in LDH and ASAT were significant and exceeded the guideline allowances. None of the other 43 analytes were affected. The changes in LDH, ASAT and K were correlated with hemolysis, which also increased with increasing transport speed.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageSpecimen transport duration/condition
    1.5 m/s
    2.0 m/s
    2.5 m/s
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system