Is the effect of hemolysis on plasma ammonia measurement overrated?
El-Khoury JM, Bunch DR, Wang S
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of the method of obtaining hemolysate, the amount of hemolysis, and storage on plasma ammonium levels. Hemolysate produced by the osmotic method was frozen and thawed and then diluted to reach the desired levels of hemolysis before being added to plasma. Different levels of hemolysis were obtained using the shearing method by passing whole blood through a needle 0, 1, 2, 5, 8 or 15 times before centrifugation.
Summary of Findings
Ammonium levels in plasma increased with increasing hemolysis. Higher levels of ammonium were measured in specimens when the added hemolysate was produced by shearing rather than by the osmotic method (with a freeze-thaw cycle), however, the rate of increase in ammonium levels with increasing hemolysis was greater when the hemolysate was produced by the osmotic method (with a freeze-thaw cycle) than by shearing. Including a freeze-thaw cycle for hemolysate produced by the osmotic method resulted in higher plasma ammonium levels than when hemolysate was produced without a freeze-thaw cycle. Ammonium levels were stable in unhemolyzed plasma left in contact with the cells for 1 h at room temperature or for 22 h when plasma was separated from cells and frozen. When plasma was stored in contact with the cells for 5 h, a similar increase in ammonium was observed in hemolyzed (by shearing) and unhemolyzed specimens. When plasma that was hemolyzed by shearing was stored in the absence of cells, the increases in ammonium were smaller than when plasma was stored in contact with the cells.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 133 ug/dL ammonium39 ug/dL ammoniumHemolysis index of 116.5Hemolysis index of 219.7Hemolysis index of 27.6Hemolysis index of 814.2Biospecimen Aliquots and Components Hemolysis AbsentDistilled water-inducedFine needle aspiration-inducedFreeze/thaw-inducedHemolysate addedStorage Freeze/thaw cycling 0 cycles1 cycleStorage Time at room temperature 0 h1 h3 h5 h