Effect of hemolysis on the concentration of insulin in serum determined by RIA and IRMA.
Chevenne D, Letailleur A, Trivin F, Porquet D
Purpose
The purpose of this paper was to determine the effects of hemolysis, storage temperature, and insulin-degrading enzyme (IDE) inhibitors on insulin in serum measured by radioimmunoassay (RIA) and immunoradiometric assay (IRMA).
Conclusion
Insulin levels, as measured by IRMA and RIA, decreased with increasing levels of hemolysate added to the serum and with increasing storage temperature. p-chloromercuriphenylsulfonic acid (pcMPS) or diamide partially attenuated the effects of added hemolysate on serum insulin levels. Slightly greater decreases in insulin were measured by IRMA than by RIA.
Studies (1)
Purpose
The purpose of this study was to determine the effects of increasing levels of hemolysis, storage temperature, and the addition of IDE inhibitors on levels of insulin in serum measured by RIA and IRMA. Red blood cells lysed by freezing were added to serum to achieve the desired concentration of hemoglobin. Details on the collection of blood for this study were not discussed.
Summary of Findings
Insulin levels, as measured by IRMA and RIA, decreased with increasing levels of hemolysate added to the serum (as much as 55-67% at 20 degrees C). Larger decreases were measured at higher storage temperatures such that after 1 h at 37 degrees C, serum with 6 g/L hemoglobin added only had 2.5% as much insulin as the original serum. The addition of pcMPS or diamide limited the hemolysate-induced decrease to <20%, but insulin levels were still significantly lower than in serum to which no hemolysate was added. The authors report that the addition of EDTA or 1.10-phenanthroline had no effect on insulin levels. Slightly greater decreases in insulin were measured by IRMA than by RIA.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinImmunoradiometric assayProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 0.4 mmol/L pcMPS0.5 g/ L hemoglobin1 g/ L hemoglobin1.10-phenanthroline2 g/ L hemoglobin4 g/ L hemoglobin5 mmol/L EDTA5 mmol/L diamide6 g/ L hemoglobinBiospecimen Aliquots and Components Hemolysis AbsentFreeze/thaw-inducedHemolysate addedStorage Storage duration 0 h1 hStorage Storage temperature 20 degrees C37 degrees C4 degrees CRadioimmunoassay Specific Technology platform IRMA