Assessing pneumatic tube systems with patient-specific populations and laboratory-derived criteria.

Kavsak PA, Mansour M, Wang L, Campeau S, Clark L, Brooks D, Trus M

Publication: Clin Chem (2012) • Vol 58 • Page 792-5PMID: 22291142View on PubMed

Purpose

The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport and conditions during PTS transport on the levels of aspartate aminotransferase, potassium, lactate dehydrogenase (LD) and phosphate in pooled blood specimens from 4 patient groups.

Conclusion

PTS transport led to increased aspartate aminotransferase, potassium and LD in some specimens, but changes depended on blood source and transport conditions. Use of a foam liner and stopping the tube without impact generally reduced the effects of PTS transport on potassium and aspartate aminotransferase and, in some cases, reduced the effect on LD. The effects of PTS transport on analytes were generally largest in blood specimens from the hematology/oncology ward and smallest in blood specimens from the intensive care unit.

Studies (1)

Purpose

The purpose of this study was to determine the effects of PTS transport and conditions during PTS transport on the levels of aspartate aminotransferase, potassium, LD and phosphate in pooled blood specimens from 4 patient groups. Leftover lithium heparin blood from 5 patients was resuspended and pooled for each location and then transported from the same substation. Transport of the reference specimen not sent by PTS is unspecified and assumed to be hand-delivery. Storage conditions and temperature of the blood prior to the study was unspecified.

Summary of Findings

PTS transport led to increased aspartate aminotransferase, potassium and LD in some specimens, but changes depended on blood source and transport conditions. Use of a foam liner and stopping the tube without impact generally reduced the effects of PTS transport on potassium and aspartate aminotransferase and, in some cases, reduced the effect on LD. The effects of PTS transport on analytes were generally largest in blood specimens from hematology/oncology ward and smallest in blood specimens from the intensive care unit.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Neoplastic
    Other diagnoses
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Emergency department patient
    Hematology/Oncology patient
    Intensive care unit patient
    Surgical oncology and orthopedics patient
    StorageSpecimen transport duration/condition
    Foam liner and tube stopped without impact
    Foam liner, tube stopped on impact
    No liner tube stopped on impact
    Tube stopped without impact
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system