Skip to main content

The stability of blood, plasma and serum constituents during simulated transport.

Felding P, Petersen PH, Hørder M

Publication: Scand J Clin Lab Invest (1981) • Vol 41 • Page 35-40PMID: 7256191View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage for 4 days at 20, 4, and -20 degrees C, conditions prior to storage, and an additional thawing of specimens on levels of clinical chemistry analytes in serum, coagulation factors in plasma, and cell counts in whole blood.

Conclusion

Serum iron, aspartate aminotransferase (ASAT), bilirubin, alkaline phosphatase (ALP), creatinine kinase, plasma coagulation factors, whole blood cell counts, hemoglobin and hematocrit were affected by storage temperature, shaking, rotation, heating and additional freeze-thaw cycling. All other analytes remained within one coefficient of variance (CV) of the assay of initial levels after all treatments.

Studies (2)

Purpose

The purpose of this study was to determine the effects of storage for 4 days at 20, 4, and -20 degrees C, rotation for 5 h, shaking for 1 h or heating to 50 degrees C for 1 h prior to storage, and an additional thawing of specimens stored at -20 degrees C on levels of clinical chemistry analytes in serum, glucose and coagulation factors in plasma, and hematology parameters in blood. Specimens for all analytes except glucose were stabilized with sodium citrate, but for glucose, plasma was stabilized with heparin sodium fluoride.

Summary of Findings

Storage for 4 days at 20 degrees C led to a 10% decrease in serum iron levels, a 16% increase in plasma coagulation factors II, VII and X by the OWREN analyzer and an unspecified increase in plasma coagulation factors II, V, VIII, IX and XII when determined by ACT levels relative to initial levels. Storage for 4 days at 4 degrees C led to an 8.7% decrease in serum iron levels and a 13% increase in plasma coagulation factors by ACT relative to initial levels. After 4 days at -20 degrees C, levels of iron in serum were increased 6.8%. When there was an additional thawing for 1 h at room temperature during the 4 days of storage at -20 degrees C, all analyte levels remained within one CV of initial levels. When serum was subjected to 5 h of rotation at room temperature prior to 4 days of storage at -20 degrees C, levels of iron, ASAT, and bilirubin in serum increased 21%, 36%, and 8.8%, respectively, and plasma coagulation factors measured by OWREN decreased 25% relative to initial levels. However, when serum was subjected to 1 h of shaking prior to storage at -20 degrees C, levels of all analytes remained within one CV of initial levels. When serum was heated to 50 degrees C for 1 h and then stored for 4 days at -20 degrees C, ALP, creatinine kinase and plasma coagulation factors (measured by OWREN) decreased by 36%, 69% and 52%, respectively. All other analytes remained within the CV of initial values, regardless of storage temperature or pre-storage treatments.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)FrozenOther Preservative
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    GlycoproteinRadioimmunoassay
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    PeptideRadioimmunoassay
    CarbohydrateEnzyme assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    No shaking or rotation
    With rotation at 40 rpm for 5 h
    With shaking at 10 Hz for 1 h
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    Refrigeration
    StorageFreeze/thaw cycling
    1 cycle
    2 cycles
    StorageStorage temperature
    -20 degrees C
    20 degrees C
    4 degrees C
    50 degrees C