Skip to main content

Effect of serum-clot contact time on clinical chemistry laboratory results.

Zhang DJ, Elswick RK, Miller WG, Bailey JL

Publication: Clin Chem (1998) • Vol 44 • Page 1325-33PMID: 9625060View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage of serum in contact with the clot on 63 clinical chemistry analytes and to determine the effects of pneumatic tube system (PTS) transport of serum in full or half-full tubes on lactate dehydrogenase (LD).

Conclusion

Storage of serum in contact with the clot affected glucose and potassium levels after 3 h, albumin and inorganic phosphorus after 6 h, and bicarbonate, chloride, C-peptide, iron, high density lipoprotein (HDL), low density lipoprotein (LDL), and total protein after 12 h. Potassium levels were less affected by storage of the serum in contact with the clot at room temperature than 32 degrees C, but significant effects were still observed after only 3 h. LD was less affected by PTS transport of specimens in full tubes than in half-filled tubes.

Studies (3)

Purpose

The purpose of this study was to determine the effects of delayed separation from the clot on the levels of 63 clinical chemistry analytes in serum. Specimens from at least 4 individuals were used for each analyte. After centrifugation, serum was refrigerated at 4 degrees C until analysis and protected from light when necessary.

Summary of Findings

Storing serum on the clot for 3 hours at 32 degrees C resulted in decreased levels of glucose that were clinically significant, while levels of albumin and inorganic phosphorus did not show clinically significant increases until serum was stored for 6 h on the clot. Levels of bicarbonate, chloride, C-peptide, iron, HDL, LDL, and total protein remained within clinically acceptable levels after storage of the serum in contact with the clot for 6 h, but between 6-24 h, iron and total protein increased and bicarbonate, chloride, C-peptide, HDL and LDL decreased. Levels of 50 other clinical chemistry analytes were not clinically affected by storage of serum in contact with the clot at 32 degrees C for 24 hours.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    LipoproteinClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    12 h
    24 h
    3 h
    6 h
    <30 min