Skip to main content

The effect of time between venipuncture, processing and freezing on the measurement of coagulation factor levels.

Engbers MJ, Cushman M, Rosendaal FR, Van Hylckama Vlieg A

Publication: J Thromb Haemost (2012) • Vol 10 • Page 1691-3PMID: 22632130View on PubMed

Purpose

The purpose of this paper was to determine the effects of storing blood at different temperatures, prior to centrifugation, on measured levels of coagulation factors.

Conclusion

None of the analytes were affected by a 2.5 h room temperature delay prior to centrifugation, but fibrinogen levels increased and factor (F) VIII levels were decreased when blood specimens were stored on ice for 2.5 h prior to centrifugation compared to those centrifuged within one hour of collection. Levels of prothrombin, antithrombin, factor (F) VIII, FIX, and FXI were unaffected by delaying centrifugation by 2.5 h, regardless of storage temperature.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storing blood for 2.5 h at room temperature or on ice, prior to centrifugation, on measured levels of coagulation factors. Sodium citrate blood was collected from the first venipuncture from 10 patients with a maximum of 25 s tourniquet time. After centrifugation plasma was stored at -80 degrees C until analysis.

Summary of Findings

None of the analytes were affected by a 2.5 h room temperature delay in centrifugation, but fibrinogen levels increased and factor VIII levels were decreased when specimens were stored on ice for 2.5 h prior to centrifugation compared to those centrifuged within the hour. Levels of prothrombin, antithrombin, FVIII, FIX, and FXI were unaffected by delaying centrifugation by 2.5 h, regardless of temperature. The authors report that the changes were similar for individuals with high and low coagulation factor levels.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    GlycoproteinHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    High levels coagulation factors
    Low levels of coagulation factors
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 h
    2.5 h
    StorageStorage temperature
    On ice (~4 degrees C)
    Room temperature (21 degrees C)