Stability of NT-proBNP in serum specimens collected in Becton Dickinson Vacutainer (SST) tubes.
Dasgupta A, Chow L, Tso G, Nazareno L
Purpose
The purpose of this paper was to determine the effects of storing serum in serum separator tubes (SST) or plain tubes at room temperature and then 4 degrees C on levels of N-terminal fragment proB-type natriuretic peptide (NT-proBNP).
Conclusion
NT-proBNP levels were not significantly affected by tube type or up to 48 h of storage.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storing serum from healthy volunteers in SST or plain tubes or storing serum from patients with suspected cardiac problems in SST tubes on NT-proBNP levels. Specimens were stored at room temperature for the first 5 h and then transferred to 4 degrees C.
Summary of Findings
NT-proBNP levels were not significantly affected by storing serum in SST or plain tubes for up to 48 h.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalCardiovascular DiseasePlatforms
PeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Plain tubeSSTStorage Storage duration 2 h24 h48 h5 h<1 hStorage Type of storage container Plain tubeSST