Skip to main content

Analytical performance of a highly sensitive C-reactive protein-based immunoassay and the effects of laboratory variables on levels of protein in blood.

Aziz N, Fahey JL, Detels R, Butch AW

Publication: Clin Diagn Lab Immunol (2003) • Vol 10 • Page 652-7PMID: 12853400View on PubMed

Purpose

The purpose of this paper was to determine the effects of patient gender, delayed centrifugation, storage temperature and freeze-thaw cycling of serum on C-reactive protein (CRP) levels. Levels were also compared between serum and plasma.

Conclusion

Patient gender, a 6 h delay in centrifugation, storage temperatures ranging from -70 degrees to room temperature and up to 7 freeze-thaw cycles did not significantly affect CRP levels in serum. Further, neither measuring levels in plasma (heparinated or EDTA) rather than serum nor measuring levels with the automated BNII nephelometer instead of an immunoassay affected CRP levels.

Studies (1)

Purpose

The purpose of this paper was to determine the effects of blood component, patient gender, delayed centrifugation, storage temperature and freeze-thaw cycling of serum on CRP levels in healthy individuals and those tested positive for human immunodeficiency virus (HIV).

Summary of Findings

CRP levels were not significantly different between EDTA-plasma, heparinated-plasma and serum or between males and females. When specimens were held at room temperature for 6 h instead of 1 h, prior to centrifugation, serum CRP levels were slightly lower, but the change was not significant. After 3 weeks of storage, serum levels of CRP in specimens stored at room temperature or 4 degrees C were not significantly different from those in specimens stored at -70 degrees C. While no statistically significant changes in CRP were noted after as many as 10 freeze-thaw cycles, some individual specimens had much lower CRP levels after 10 cycles leading the authors to conclude that specimens should not be subjected to more than 7 freeze-thaw cycles. There was a very strong correlation between CRP levels measured by the immunoassay and those measured using the automated BNII nephelometer.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    AIDS/HIV-related
    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    ProteinImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Heparin
    None
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageFreeze/thaw cycling
    1 cycle
    10 cycles
    3 cycles
    7 cycles
    StorageStorage temperature
    -70 degrees C
    4 degrees C
    Room temperature
    StorageTime at room temperature
    1 h
    6 h
    Immunoassay SpecificTechnology platform
    BNII nephelometer