Skip to main content

Quantitative miRNA expression analysis using fluidigm microfluidics dynamic arrays.

Jang JS, Simon VA, Feddersen RM, Rakhshan F, Schultz DA, Zschunke MA, Lingle WL, Kolbert CP, Jen J

Publication: BMC Genomics (2011) • Vol 12 • Page 144PMID: 21388556View on PubMed

Purpose

The purpose of this paper was to determine the effects of using formalin-fixed paraffin-embedded (FFPE) lung specimens instead of frozen lung specimens on microRNA (miRNA) profiles measured using conventional and dynamic array real-time PCR assays and microarrays.

Conclusion

The real-time RT-PCR cycle threshold (CT) values were higher when measured using the ABI 7900 rather than the Fluidigm dynamic array. CT values were also lower for frozen specimens than for the corresponding FFPE specimens, but the expression in frozen and FFPE specimens was highly correlated (R2=0.87-0.98). Further, there was a 0.94 correlation between frozen and FFPE specimens using microarray. The range in fold change in expression between frozen and FFPE specimens was larger (0-13 fold) using the Fluidigm dynamic arrays than using microarray (0-3.5 fold).

Studies (1)

Purpose

The purpose of this study was to determine the effects of using FFPE lung specimens instead of frozen lung specimens on microRNA (miRNA) profiles measured using conventional and dynamic real-time PCR assays and microarrays. miRNA was extracted from xylene deparaffinized 10 uM sections of FFPE tissue using miRNeasy or RecoverAll. miRNA was extracted from frozen specimens using miRNeasy. miRNA was preamplified with pooled TaqMan assays.

Summary of Findings

When real-time RT-PCR was conducted using 48.48 Fluidigm dynamic arrays, the average CT was only 12.48 compared to 16.21 using the ABI7900. The CT values were lower for frozen specimens than for the corresponding FFPE specimens, regardless of machine, but the data generated was highly correlated (R2=0.87-0.98). A linear relationship between CT and input quantity was observed for FFPE and frozen specimens using the Fluidigm dynamic arrays. 33-35% of miRNAs were detected on the arrays and there was a correlation of 0.94 between frozen and FFPE specimens. The range in fold change in expression between frozen and FFPE specimens was larger (0-13 fold) using the Fluidigm dynamic arrays than using microarray (0-3.5 fold).

  • Biospecimens

    TissueLung
  • Preservation Types

    FormalinFrozen
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    RNADNA microarray
    RNAReal-time qRT-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Formalin (buffered)
    Frozen
    Real-time qRT-PCR SpecificTargeted nucleic acid
    7900 HT
    Fluidigm dynamic array
    Real-time qRT-PCR SpecificTemplate/input amount
    10 ng
    100 ng
    25 ng
    50 ng