Effects of time and temperature during attachment of sections to microscope slides on immunohistochemical detection of antigens.
Jones WT, Stockard CR, Grizzle WE
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of time and temperature during tissue section attachment to slides on IHC staining using a panel of 11 different antibodies and 9 neoplastic and benign tissues including colon and other unspecified tissues. Plus glass microscope slides were used.
Summary of Findings
Attachment times of 1, 4, or 24 h did not affect immunostaining intensity of B-cell lymphoma 2 (Bcl-2), and the authors state that generally, immunostaining intensities of the other tested antigens were also unaffected by this range of slide attachment times. There were no differences in the intensity of proliferating cell nuclear antigen (PCNA) immunostaining with slide attachment temperatures of 58 or 68 degrees C (16 h attachment). However, a slide attachment temperature of 80 degrees C for 16 h tended to decrease immunostaining intensity for PCNA and, the authors state, all other antigens tested.
Biospecimens
TissueColorectalTissueOtherPreservation Types
FormalinDiagnoses
NeoplasticNeoplasticBenignPlatforms
ProteinImmunohistochemistryMorphologyLight microscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Tissue section adhesion 1 h16 h24 h4 h58 degrees C68 degrees C80 degrees C