Stability of serum thyroid hormones following 8-11 years of cold storage.

Panesar NS, Lit LC

Publication: Clin Chem Lab Med (2010) • Vol 48 • Page 409-12PMID: 20014958View on PubMed

Purpose

The purpose of this paper was to determine the effects of 8-11 years of frozen storage and type of assay on measured levels of thyroid stimulating hormone (TSH), free thyroxine (FT4) and free triiodothyronine (FT3) in serum.

Conclusion

Regardless of the type of assay used, TSH levels were slightly lower and FT4 and FT3 levels were higher after 8-11 years of frozen storage than when measured in fresh serum using the ACS180. Levels of FT4 and FT3 were dependent on the analyzer, but TSH levels were similar when using the Centaur, E170, or Architect assays.

Studies (1)

Purpose

The purpose of this study was to determine the effects of 8-11 years of frozen storage and type of assay used for measuring TSH, FT4 and FT3 in serum from pregnant women. The ACS180 assay was used only for fresh specimens, and a subset of specimens were assayed fresh using the E170 assay. Frozen specimens were stored at -80 degrees C for 8-11 years or were newly archived at -80 degrees C (unspecified duration).

Summary of Findings

Serum FT4 levels were 17.6% higher in specimens stored frozen for 8-11 years and measured using the E170 than when measured using ACS180 in the same specimens prior to freezing, but only an 11% difference was found when fresh specimens were measured using ACS180 and E170. In specimens stored for 8-11 years, FT3 levels measured using the E170 were 42% higher than when measured in the same specimens prior to freezing with the ACS180, but when fresh specimens were measured with the E170, FT3 levels were 10% lower than when measured with the ACS180. In specimens stored frozen 8-11 years, TSH levels were slightly lower using all three assays than when measured in the same specimens prior to freezing using the ACS180. TSH results showed significant correlations between the 4 assays in specimens stored 8-11 years, newly archived specimens and fresh specimens (r2>0.96 for all). FT4 levels had correlation coefficients of 0.8-0.93 between measurements with the different assays in fresh and newly archived specimens. FT3 levels had correlation coefficients between methods of 0.59-0.89 in fresh and newly archived specimens.

  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)Frozen
  • Diagnoses

    Pregnant
  • Platforms

    GlycoproteinClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Pregnant (5-41 weeks)
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    StorageStorage duration
    0 days
    0 years
    8-11 years
    Clinical chemistry/auto analyzer SpecificTechnology platform
    ACS180
    Architect
    Centaur
    E170