Determining stability of stored samples retrospectively: the validation of glycated albumin.

Nathan DM, Steffes MW, Sun W, Rynders GP, Lachin JM

Publication: Clin Chem (2011) • Vol 57 • Page 286-90PMID: 21030684View on PubMed

Purpose

The purpose of this paper was to determine the effects of long-term storage of EDTA-whole blood at -70 degrees C on plasma levels of glycated albumin (GA).

Conclusion

Similar correlations between GA levels after storage and glycohemoglobin (HbAIc) levels prior to storage were found in specimens stored for <2 years (exact duration unspecified) and those stored for 19-23 years; however, GA was significantly higher in specimens stored for 19-23 years than in specimens stored for <2 years.

Studies (1)

Purpose

The purpose of this study was to determine the effects of long-term storage of EDTA-whole blood at -70 degrees C on plasma levels of GA. The authors do not specify the duration of storage of specimens stored for <2 years, but state the specimens were collected in 2009. HbAIc was measured in fresh specimens using the same method for the entire study duration.

Summary of Findings

Similar correlations between GA levels in stored specimens and HbAIc levels in fresh specimens were found in specimens stored for <2 years (r2=0.592) and those stored for 19-23 years (r2=0.696), and the slope of the linear regression was similar (0.198 versus 0.187). However, GA was significantly higher in specimens stored for 19-23 years than in specimens stored for <2 years (p=0.0007).

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Diabetes Type 1
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    ProteinHPLC
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    19-23 years
    <2 years