Effect of Heparin Contaminated with Oversulfated Chondroitin Sulfate on the Collection and Analysis of Plasma

Bosworth T, Chan Y, Nguyen T, Gutman S, Gutierrez A, Kalush F, Gwise T, Pollack S, Guo J, Patwardhan D, Buhse L, Reepmeyer JC, Remaley AT

Publication: The Open Clinical Chemistry Journal (2010) • Vol 3 • Page 5-7

Purpose

The purpose of this paper was to determine the effects of oversulfated chondroitin sulfate (OSCS) in heparin blood collection tubes on the formation of microclots and the measurement of 29 common clinical chemistry analytes in plasma.

Conclusion

Specimens with increasing amounts of OSCS had increasing incidences of microclots, while none were observed in the 50 specimens containing no OSCS. 6 out of 29 clinical chemistry and immunoassay tests performed yielded significantly different results in the presence of 20% OSCS compared to specimens with no OSCS. Importantly, these biases did not exceed the medically allowable error rates.

Studies (1)

Purpose

The purpose of this study was to determine the effects of OSCS in heparin blood collection tubes on the formation of microclots and the measurement of 29 common clinical chemistry analytes in plasma. Specimens were collected into plastic syringes with no anticoagulants and transferred immediately into glass tubes containing various concentrations of OSCS relative to heparin representative of a conceivable range of OSCS contamination.

Summary of Findings

Specimens with increasing amounts of OSCS had increasing incidences of microclots (18-76%), while none were observed in the 50 specimens containing no OSCS. 6 out of 29 clinical chemistry and immunoassay tests performed (lactate dehydrogenase, triiodothyronine, potassium, total protein, chloride, and uric acid) yielded significantly different results, ranging from average biases of -5.7% to +18%, in the presence of the highest concentration of OSCS (20%), compared to specimens with no OSCS. Importantly, these biases did not exceed the medically allowable error rates. pH was among the analytes unaffected by OSCS.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    MorphologyMacroscopic observation
    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    GlycoproteinImmunoassay
    SteroidImmunoassay
    PeptideImmunoassay
    ProteinImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Heparin
    Multiple concentrations evaluated
    Biospecimen Aliquots and ComponentsBiospecimen components
    0% OSCS
    10% OSCS
    20% OSCS
    5% OSCS