Effect of sample storage on quantitation of lipoprotein(a) by an enzyme-linked immunosorbent assay.
Evans RW, Sankey SS, Hauth BA, Sutton-Tyrrell K, Kamboh MI, Kuller LH
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage duration and temperature on Lp(a) levels in serum. Specimens were collected from 25 healthy volunteers, in serum separator tubes, after an overnight fast, and aliquoted into polypropylene tubes prior to freezing.
Summary of Findings
Generally, levels of Lp(a) declined with increasing storage duration, becoming significant at 6 months after storage at either -20 or -70 degrees C, but larger declines were found in specimens stored at -20 degrees C than -70 degrees C (p<0.05). Interestingly, Lp(a) levels were higher after 3 months at -20 or -70 degrees C than in fresh specimens, which the authors attributed to analytical variability. Although Lp(a) declined by 36.9% and 19.1% during the three years of storage at -20 and -70 degrees C, respectively, a strong correlation with baseline levels was maintained (r=0.970 and r=0.987, respectively). The effects of storage were less when the baseline Lp(a) was > 8.0 mg/dL, and the authors report that large Apolipoprotein(a) isoforms were more likely to be lost than small isoforms at both temperatures.
Biospecimens
FluidSerumPreservation Types
FrozenNone (Fresh)Diagnoses
NormalPlatforms
LipoproteinELISAPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level < 8 mg/dL> 8 mg/dLBiospecimen Preservation Type of fixation/preservation FrozenNone (fresh)Storage Storage duration 0 days1 week1 year2 years3 months3 years4 weeks6 monthsStorage Storage temperature -20 degrees C-70 degrees C