A comparative study of plastic (polypropylene) and glass syringes in blood-gas analysis.
Evers W, Racz GB, Levy AA
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of storing blood in plastic versus glass syringes and refrigerated storage duration on PO2, PCO2, and pH of blood specimens. All specimens were initially collected in a heparinized syringe and aliquoted into either plastic or glass sample syringes via a sealed multiconnector.
Summary of Findings
No statistically significant changes in blood pH were seen after 4 hours of refrigerated storage in either plastic or glass syringes. PCO2 values fell during the first 30 min of storage at 4 degrees C and thereafter, rose steadily. PO2 values also fell during the first 30 min of storage at 4 degrees C, but no further changes were observed. There were no significant differences in PCO2 or PO2 values in specimens stored in plastic rather than glass syringes.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Other PreservativeDiagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation None (fresh)RefrigerationStorage Storage duration 0 min120 min240 min30 min60 minStorage Type of storage container Glass syringePlastic syringe