The robustness of the Roche 2nd generation homogenous HDL cholesterol (PEGME) method: assessment of the effect of serum sample storage for up to 8 years at -80 degrees C.

Charlton-Menys V, Liu Y, Moorhouse A, Durrington PN

Publication: Clin Chim Acta (2007) • Vol 382 • Page 142-4PMID: 17511978View on PubMed

Purpose

The purpose of this paper was to determine the effects of frozen storage on measurement of cholesterol in serum.

Conclusion

After normalizing measurements taken on two different instruments by using the center for disease control (CDC) method, a small (<10%) but significant decline in average high density lipoprotein cholesterol (HDL-C) levels was observed after storage. Altman-Bland analysis revealed that the effect of storage was concentration-dependent but independent of storage duration.

Studies (1)

Purpose

The purpose of this study was to determine the effects of concentration, frozen storage and analysis method on measurement of cholesterol in serum. Specimens were analyzed on the day of collection using the CHOP-PAP method with a Cobas Mira analyzer and then analyzed 5-8 years later with the Roche 2'nd generation direct homogeneous PEGME method. The CDC method was used to normalize the results using the two methods.

Summary of Findings

Using fresh specimens, the Roche method measured HDL-C levels that were 7.9% lower than the CDC method, while the CHOP-PAP method measured HDL-C levels that were 13.9% higher than the CDC method. After alignment to the CDC values, the Roche method measured HDL-C levels in fresh specimens that were 3.6% lower than the CHOP-PAP method (p<0.001), and these results were highly correlated (r=0.896). After storage, there was a small (<10%) but significant decline in normalized HDL-C levels (p<0.001). Altman-Bland analysis revealed that the effect of storage was concentration-dependent (p<0.0001) but independent of storage duration.

  • Biospecimens

    FluidSerum
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    StorageStorage duration
    0 days
    5-8 years
    Clinical chemistry/auto analyzer SpecificTechnology platform
    CDC clinical chemistry standardization section method
    CHOP-PAP method with a Cobas Mira analyzer
    Roche 2'nd generation direct homogeneous PEGME method