Skip to main content

To mix or not to mix venous blood samples collected in vacuum tubes?

Parenmark A, Landberg E

Publication: Clin Chem Lab Med (2011) • Vol 49 • Page 2061-3PMID: 21899494View on PubMed

Purpose

The purpose of this paper was to determine the effects of mixing strategy after collection of blood on clinical chemistry and hematology analytes.

Conclusion

Activated partial thromboplastin time (APTT), aspartate aminotransferase (AST), bilirubin, creatinine kinase (CK), C-reactive protein (CRP), iron, potassium, free thyroxine (fT4), thyroid stimulating hormone (TSH), hematocrit, mean corpuscular volume (MCV), and hemoglobin levels, and leukocyte, platelet, and erythrocyte counts were not affected by mixing strategy. However, prothrombin time (PT), lactate dehydrogenase (LD) levels, and sodium levels were. Hemolysis index was increased in immediately mixed specimens compared to unmixed specimens in tubes containing separating gel and lithium heparin and in tubes containing citrate.

Studies (1)

Purpose

The purpose of this study was to determine the effects of mixing strategy after collection of blood on clinical chemistry and hematology analytes. Mixing was performed using a horizontal mixing tray for 1 min (15 inversions). After mixing (or not), all tubes were placed upright in a rack for 30 min prior to centrifugation.

Summary of Findings

APTT, AST, bilirubin, CK, CRP, iron, potassium, fT4, TSH, hematocrit, MCV, and hemoglobin levels, and leukocyte, platelet, and erythrocyte counts were not affected by mixing strategy. However, PT showed a small but significant increase (average 2%, p<0.05) when tubes were not mixed compared to those that were mixed immediately or after 5 min. Further, plasma LD levels showed small but significant decreases (median 4.1%, p<0.05) when tubes were not mixed compared to those that were mixed immediately or after 5 min. In addition, sodium levels showed small but significant decreases (median 1%, p<0.05) in tubes that were not mixed or subjected to delayed mixing compared to those that were mixed immediately. Hemolysis index was increased in immediately mixed specimens compared to unmixed specimens in tubes containing separating gel and lithium heparin, and in tubes containing citrate.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    MorphologyHematology/ auto analyzer
    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    Cell count/volumeSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Citrate
    Heparin
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    5 min horizontal rest followed by inversion 15 times
    6 min horizontal rest with no mixing
    Immediately inverted 15 times