Evaluation of Maxwell(®) 16 for automated DNA extraction from whole blood and formalin-fixed paraffin embedded (FFPE) tissue.
Khokhar SK, Mitui M, Leos NK, Rogers BB, Park JY
Purpose
The purpose of this paper was to determine the effects of DNA isolation method on yield and amplification success from EDTA-blood and formalin-fixed paraffin-embedded (FFPE) liver sections.
Conclusion
The highest DNA yield was obtained from blood extracted with the Maxwell 16 LEV blood DNA kit and from FFPE liver sections extracted with Qiagen AllPrep FFPE DNA/RNA kit. Beta-actin and a 286 bp fragment of CDKL5 were successfully amplified using DNA from 10 blood and 10 FFPE liver specimens, respectively, regardless of extraction method. While 402 bp fragement of exon 16 of CDKL5 was successfully amplified in all 10 specimens extracted with the Maxwell 16 FFPE tissue LEV DNA kit, it was not amplified in all 10 specimens extracted by the other methods. A 618 bp fragment of exon 5 was amplified in 2 of 10 FFPE sections extracted with Qiagen AllPrep FFPE DNA/RNA kit, but the same fragment was not amplified in any FFPE sections extracted with Maxwell 16 FFPE tissue LEV DNA kit or the Qiagen AllPrep DNA/RNA kit.
Studies (2)
Purpose
The purpose of this study was to determine the effects of DNA isolation method on yield and amplificability of DNA from EDTA-blood.
Summary of Findings
A significantly higher average DNA yield was obtained from blood extracted with the Maxwell 16 LEV blood DNA kit (118.68 ng/ul) than with the NucliSens EasyMag Blood DNA kit (21.33 ng/ul, p<0.0001) or the QIAamp DNA Blood mini kit (45.69 ng/ul, p<0.0001). The A260/280 ratio was 1.84 for specimens extracted with Maxwell 16 LEV blood DNA kit and the QIAamp DNA Blood mini kit, but it was only 1.58 when DNA was extracted using NucliSens EasyMag Blood DNA kit. Beta-actin was successfully amplified from all 10 blood specimens, regardless of extraction method.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
DNAPCRDNASpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method Maxwell 16 LEV blood DNA kitNucliSens EasyMag Blood DNA kitQIAamp DNA Blood mini kitPCR Specific Targeted nucleic acid Beta-actin
Purpose
The purpose of this study was to determine the effects of DNA isolation method and amplicon size on PCR success using FFPE liver specimens. DNA was extracted from 20 micron sections.
Summary of Findings
Significantly higher DNA yields were obtained from FFPE sections extracted using the Qiagen AllPrep FFPE DNA/RNA kit (129 ng/ul) than the Qiagen AllPrep DNA/RNA kit (7.9 ng/ul, p<0.0001) or the Maxwell 16 FFPE tissue LEV DNA kit (17.42, p<0.0001). The A260/280 ratio was 1.9 or greater for DNA extracted from FFPE liver sections using each of the three DNA extraction kits. A 286 bp fragment of CDKL5 exon 3 was successfully amplified using DNA from all 10 FFPE liver sections, regardless of extraction method, but a while 402 bp fragment of exon 16 of CDKL5 was successfully amplified in all 10 specimens extracted with the Maxwell 16 FFPE tissue LEV DNA kit, it was not amplified in all 10 specimens extracted by the other methods. In contrast, a 618 bp fragment of exon 5 was amplified in 2 of 10 FFPE sections extracted with Qiagen AllPrep FFPE DNA/RNA kit, but the same fragment was not amplified in DNA from FFPE sections extracted with Maxwell 16 FFPE tissue LEV DNA kit or the Qiagen AllPrep DNA/RNA kit.
Biospecimens
TissueLiverPreservation Types
FormalinDiagnoses
Not specifiedPlatforms
DNAPCRDNASpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method Maxwell 16 FFPE tissue LEV DNA kitQiagen AllPrep DNA/RNA kitQiagen AllPrep FFPE DNA/RNA kitPCR Specific Length of gene fragment 286 bp402 bp618 bpPCR Specific Targeted nucleic acid CDKL5 exon 16CDKL5 exon 3CDKL5 exon 5