Skip to main content

Influence of the needle bore size on platelet count and routine coagulation testing.

Lippi G, Salvagno GL, Montagnana M, Poli G, Guidi GC

Publication: Blood Coagul Fibrinolysis (2006) • Vol 17 • Page 557-61PMID: 16988551View on PubMed

Purpose

The purpose of this paper was to determine the effects of needle gauge on platelet counts and routine coagulation testing.

Conclusion

With specimens collected using a 21-gauge needle used as a reference, D-dimer levels were significantly higher when blood was collected using a 25-gauge needle, and platelet counts were significantly lower when collected using a 23-gauge or 25-gauge needle. No significant differences were seen between 21-gauge needle specimens and 23-gauge or 25-gauge needle specimens with respect to activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen, or free hemoglobin.

Studies (1)

Purpose

The purpose of this study was to determine the effects of needle gauge on platelet counts, PT, APTT, and fibrinogen, D-dimer, and free hemoglobin levels in plasma. Three sequential venipunctures, on alternating arms and different veins, were used to collect specimens within 5 min.

Summary of Findings

With specimens collected using a 21-gauge needle used as a reference, D-dimer levels were significantly higher when blood was collected using a 25-gauge needle (p<0.001), and platelet counts were significantly lower when collected using a 23-gauge or 25-gauge needle (both p<0.001). No significant differences were seen between 21-gauge needle specimens and 23-gauge or 25-gauge needle specimens with respect to APTT, PT, fibrinogen, or free hemoglobin.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    PeptideELISA
    MorphologyHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
    Cell count/volumeHematology/ auto analyzer
    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionNeedle gauge
    21-gauge
    23-gauge
    25-gauge
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    3rd collection