Pseudohyperkalaemia caused by recentrifugation of blood samples after storage in gel separator tubes.
Hira K, Ohtani Y, Rahman M, Noguchi Y, Shimbo T, Fukui T
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of recentrifugation after storage of serum in gel separator tubes on clinical chemistry and hematological analytes. Aliquots of blood from a single volunteer were centrifuged and then refrigerated at 4 degrees C for up to 72 h, at which point, half were recentrifuged before analysis.
Summary of Findings
Sodium and potassium levels were relatively stable in serum stored in serum gel tubes for 72 h after centrifugation; however, when a second centrifugation was performed immediately before analysis, decreases in sodium levels and storage time-dependent increases in potassium levels were observed. The authors report that neither storage nor recentrifugation affected white blood cell counts, hematocrit, platelet counts, urea, nitrogen, creatinine or lactate dehydrogenase levels.
Biospecimens
FluidSerumPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Different number of centrifugation steps comparedStorage Storage duration 0 h24 h48 h72 h