Influence of mechanical trauma of blood and hemolysis on PFA-100 testing.
Lippi G, Fontana R, Avanzini P, Aloe R, Ippolito L, Sandei F, Favaloro EJ
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically-induced hemolysis on HI, cell free hemoglobin, RBC counts, platelet counts, platelet aggregation, and plasma LD levels. Hemolysis was induced by passing aliquots of whole blood through a 30-gauge needle 2 (mild hemolysis) or 4 times (gross hemolysis).
Summary of Findings
Passing whole blood through a 30-gauge needle resulted in increases in HI and cell free hemoglobin. Increasing degrees of hemolysis resulted in increased levels of LD, but decreased platelet and RBC counts. 4 of 9 mildly hemolyzed aliquots (2 passages through 30-gauge needle) resulted in a flow obstruction in the PFA-100 analyzer using both collagen/adenosine diphosphate (ADP) activation and collagen/epinephrine activation, and the remaining 5 showed increased closure times compared to non-hemolyzed aliquots. 6 of 9 and 9 of 9 grossly hemolyzed specimens (4 passages through 30-gauge needle) resulted in a flow obstruction in the PFA-100 analyzer for collagen/ADP activation and collagen/epinephrine activation, respectively. The remaining 3 aliquots had increased closure times.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Cell count/volumeSpectrophotometryProteinSpectrophotometryCell count/volumeHematology/ auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 0-0.5 g/L cell free hemoglobin11.5-12.0 g/L cell free hemoglobin6.0-6.5 g/L cell free hemoglobinBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot inducedHematology/ auto analyzer Specific Reaction solution Collagen/ADPCollagen/epinephrine