Studies on in vitro hemolysis and utility of corrective formulas for reporting results on hemolyzed specimens.
Lippi G, Avanzini P, Pavesi F, Bardi M, Ippolito L, Aloe R, Favaloro EJ
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically-induced hemolysis on hematological testing, HI, and levels of glucose, LD, AST, and K in plasma. Hemolysis was induced by aspiration of whole blood aliquots through a 30-gauge needle from 1 to 5 times.
Summary of Findings
HI and cell free hemoglobin increased with increasing numbers of aspirations through a 30-gauge needle. While increasing degrees of hemolysis did not significantly affect total hemoglobin or glucose measurements, and only marginally decreased WBC counts, increasing HI resulted in significantly decreased RBC counts, hematocrit, and MCV, as well as significantly increased MCH and LD, AST, and K in plasma. Gross hemolysis (HI>100) resulted in increasd platelet counts. The authors conclude that use of corrective formulas for hemolyzed specimens may be unreliable due to variability between individuals and the dishomogenous nature of mechanical injury of blood.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerCell count/volumeSpectrophotometryProteinSpectrophotometryCell count/volumeHematology/ auto analyzerProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 14.0 g/L free hemoglobin22.0 g/L free hemoglobin29.0 g/L free hemoglobin37.0 g/L free hemoglobin7.0 g/L free hemoglobin<0.5 g/L free hemoglobinBiospecimen Aliquots and Components Blood and blood products PlasmaWhole bloodBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot induced