Methods for isolation of cell-free plasma DNA strongly affect DNA yield.

Fleischhacker M, Schmidt B, Weickmann S, Fersching DM, Leszinski GS, Siegele B, Stötzer OJ, Nagel D, Holdenrieder S

Publication: Clin Chim Acta (2011) • Vol 412 • Page 2085-8PMID: 21861994View on PubMed

Purpose

The purpose of this paper was to determine the effects of DNA isolation method, targeted gene, and prior specimen transport on cell-free DNA yields from plasma.

Conclusion

The average DNA yield was highest in specimens extracted using the MagNA Pure isolation system and lowest when extracted using QIAamp DNA blood Midi kit. Real time qPCR based DNA quantification was twice as high when Beta-globin was the targeted gene than when determined by amplification of Human endogenous virus (ERV) or Glyceraldehyde 3-phosphate dehydrogenase (GAPDH). In general, the laboratory where the DNA was extracted had no effects on DNA yield, but significantly more DNA was extracted using the MagNA PURE isolation system in the untransported specimen compared to after mailing specimens from Berlin to the Munich site.

Studies (1)

Purpose

The purpose of this study was to determine the effects of isolation method, targeted gene and prior specimen transport the on quantification of cell-free DNA yields from plasma using real-time PCR. Plasma was obtained from pulmonary and oncology patients and stored at -80 degrees C in Berlin prior to the study. Plasma aliquots from each patient were thawed, pooled, and then realiquoted for shipping to Munich on dry ice or for additional storage in Berlin at -80 degrees C.

Summary of Findings

The average DNA yield was 2-3 times higher in specimens extracted using the MagNA Pure isolation system than those extracted using NucleoSpin kit (p<0.0001), and average yields were 5-10 times higher using MagNA Pure than using QIAamp DNA blood Midi kit (p<0.0001). Real time qPCR based DNA quantification was twice as high when Beta-globin was the targeted gene than when determined by amplification of ERV or GAPDH (both p<0.001). Mailed specimens had decreased DNA yields when extracted using the MagNA PURE isolation system, and consequently, the Munich laboratory had smaller yields than the Berlin site using MagNA PURE, but not the other kits. (p=0.0126). The authors report higher DNA yields from cancer patients than from healthy controls, regardless of extraction method.
  • Biospecimens

    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    NeoplasticBenign
    NeoplasticCarcinoma
    NeoplasticLymphoma
    Normal
  • Platforms

    DNAReal-time qPCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageBetween site transportation method
    Mailed
    Not transported
    Analyte Extraction and PurificationAnalyte isolation method
    MagNA Pure isolation system
    NucleoSpin kit
    QIAamp DNA blood Midi kit
    Real-time qPCR SpecificTargeted nucleic acid
    Beta-globin
    ERV
    GAPDH