Influence of in vitro hemolysis on hematological testing on Advia 2120.
Lippi G, Musa R, Avanzini P, Aloe R, Pipitone S, Sandei F
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of hemolysis on routine hematological testing and plasma lactate dehydrogenase levels. Hemolysis was induced by passing aliquots of blood through a 30-gauge needle 5 (mild hemolysis) or 10 times (gross hemolysis).
Summary of Findings
Mechanically-induced hemolysis (both mild and gross) resulted in significant increases in mean hemoglobin content, mean corpuscular hemoglobin concentration, platelet counts, mean platelet volume, monocyte counts, basophil counts, large unstained cells, and plasma lactate dehydrogenase levels compared to specimens with no hemolysis. Further, the mechanically hemolyzed specimens showed significant decreases in red blood cell counts, hematocrit, mean corpuscular volume, and lymphocyte counts compared to non-hemolyzed specimens. Overall, no significant differences were measured in the white blood cell counts between hemolyzed and non-hemolyzed specimens, but the white blood cell differentials did change as noted above. Hemoglobin, red blood cell distribution width, neutrophil counts, and eosinophils counts were unaffected by mechanically-induced hemolysis.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Cell count/volumeHematology/ auto analyzerProteinHematology/ auto analyzerProteinClinical chemistry/auto analyzerCell count/volumeSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 0-0.5 g/L hemoglobin4.0-4.5 g/L hemoglobin9.0-9.5 g/L hemoglobinBiospecimen Aliquots and Components Blood and blood products PlasmaWhole bloodBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot induced