Preanalytical mixing of whole-blood specimens in the context of the Athlete Passport.

Ashenden M, Clarke A, Sharpe K, d'Onofrio G, Allbon G, Gore CJ

Publication: J Clin Pathol (2012) • Vol 65 • Page 8-13PMID: 22049219View on PubMed

Purpose

The purpose of this paper was to determine the effects of biospecimen mixing methods on complete blood counts (CBC).

Conclusion

No preliminary automated mixing was required for accurate CBC when the analyzer inverted tubes just prior to analysis, but in specimens inverted by hand, an additional 1 min of vertical mixing or 5 min of horizontal mixing were required to homogenize the specimens prior to analysis. Slight differences in red blood cell (RBC) counts and hemoglobin content were observed in specimens which were manually inverted compared to specimens mechanically mixed for 15 min, but allowing specimens to reach room temperature before manual inversion attenuated these effects. Mixing method and duration had little to no effects on white blood cell (WBC) counts, platelet counts, or the percentage of reticulocytes, but mean cell volume decreased with increasing duration of horizontal or vertical mixing.

Studies (1)

Purpose

The purpose of this study was to determine the effects of warming specimens prior to mixing, inversion of the specimen in the machine or by hand, and mechanical mixing method and duration on CBC results. All specimens were refrigerated and left undisturbed for 36 h before mixing.

Summary of Findings

When the analyzer used for analysis inverted each specimen 10 times just prior to analysis, prior vertical or horizontal mechanical mixing for up to 15 min had no effects on the RBC counts, hemoglobin, or percentage of reticulocytes. However, when the specimens were inverted by hand 10 times prior to analysis, the RBC counts were lower in the specimens that were not previously mixed using a mechanical mixer. After 1 min of vertical mixing or 5 min of horizontal mixing, RBC counts and hemoglobin were indistinguishable from specimens mixed vertically or horizontally for 15 min. Allowing specimens to reach room temperature before mixing increased the RBC counts and hemoglobin content in specimens that were only manually inverted before analysis and slightly decreased the RBC counts and hemoglobin content of specimens that were manually inverted and mechanically mixed for 15 min prior to analysis. Further, specimens that were allowed to warm to room temperature prior to manual inversion had slight, but significant, differences in RBC counts and hemoglobin content to specimens that were mixed mechanically for 15 min horizontally or vertically. Mixing method and duration had little to no effects on WBC counts and platelet counts, but mean cell volume decreased with increasing duration of horizontal or vertical mixing.

  • Biospecimens

    FluidBlood
  • Preservation Types

    Other Preservative
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    0 min
    0.5 min
    1 min
    10 min
    15 min
    2 min
    5 min
    At 4 degrees C
    Automated inversion
    Equilibrated to room temperature
    Horizontal mixing
    Manual inversion
    Vertical mixing
    Biospecimen Aliquots and ComponentsBlood processing method
    Inverted in the machine
    Inverted manually