Influence of hemolysis on troponin testing: studies on Beckman Coulter UniCel Dxl 800 Accu-TnI and overview of the literature.
Lippi G, Avanzini P, Dipalo M, Aloe R, Cervellin G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically induced hemolysis on Accu-TnI measurements in plasma. Specimens were separated into three aliquots, the first of which was centrifuged and analyzed, while the second and third aliquots were aspirated through a 0.5 mL syringe equipped with a 30 gauge needle 1 and 2 times, respectively, inducing hemolysis prior to centrifugation and analysis.
Summary of Findings
When unhemolyzed specimens had Accu-TnI values >0.20 ug/L, aliquots subjected to mechanically induced hemolysis by aspiration through a fine needle once or twice showed median decreases in Accu-TnI values of 10 and 17%, respectively. These decreases did not exceed the acceptable 20% variation to be viewed as clinically significant in any of the specimens aspirated once, but the decreases exceeded 20% in 3 of the 12 aliquots aspirated twice. When unhemolyzed specimens had Accu-TnI values <0.04 ug/L, mechanically induced hemolysis did not change the Accu-TnI measurements. In conclusion, Accu-TnI values decreased in hemolyzed specimens when initial values were >0.20 ug/L, but clinical significance was only reached when cell-free hemoglobin was >16.5 g/L (aliquots aspirated twice).
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerProteinSpectrophotometryCell count/volumeSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level Accu-TnI values <0.04 ug/LAccu-TnI values >0.20 ug/LBiospecimen Aliquots and Components Biospecimen components 16.5-29.5 g/L cell-free hemoglobin9.5-16.0 g/L cell-free hemoglobin<0.5 g/L cell-free hemoglobinBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot induced