Influence of hemolysis on troponin testing: studies on Beckman Coulter UniCel Dxl 800 Accu-TnI and overview of the literature.

Lippi G, Avanzini P, Dipalo M, Aloe R, Cervellin G

Publication: Clin Chem Lab Med (2011) • Vol 49 • Page 2097-2100PMID: 21892912View on PubMed

Purpose

The purpose of this paper was to determine the effects of mechanically induced hemolysis on cardiac troponin I (TnI) testing (Accu-TnI) by the Beckman Coulter UniCel Dxl 800.

Conclusion

Accu-TnI values decreased in hemolyzed specimens but clinical significance was only reached when cell-free hemoglobin was more than 16.5 g/L.

Studies (1)

Purpose

The purpose of this study was to determine the effects of mechanically induced hemolysis on Accu-TnI measurements in plasma. Specimens were separated into three aliquots, the first of which was centrifuged and analyzed, while the second and third aliquots were aspirated through a 0.5 mL syringe equipped with a 30 gauge needle 1 and 2 times, respectively, inducing hemolysis prior to centrifugation and analysis.

Summary of Findings

When unhemolyzed specimens had Accu-TnI values >0.20 ug/L, aliquots subjected to mechanically induced hemolysis by aspiration through a fine needle once or twice showed median decreases in Accu-TnI values of 10 and 17%, respectively. These decreases did not exceed the acceptable 20% variation to be viewed as clinically significant in any of the specimens aspirated once, but the decreases exceeded 20% in 3 of the 12 aliquots aspirated twice. When unhemolyzed specimens had Accu-TnI values <0.04 ug/L, mechanically induced hemolysis did not change the Accu-TnI measurements. In conclusion, Accu-TnI values decreased in hemolyzed specimens when initial values were >0.20 ug/L, but clinical significance was only reached when cell-free hemoglobin was >16.5 g/L (aliquots aspirated twice).

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinClinical chemistry/auto analyzer
    ProteinSpectrophotometry
    Cell count/volumeSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    Accu-TnI values <0.04 ug/L
    Accu-TnI values >0.20 ug/L
    Biospecimen Aliquots and ComponentsBiospecimen components
    16.5-29.5 g/L cell-free hemoglobin
    9.5-16.0 g/L cell-free hemoglobin
    <0.5 g/L cell-free hemoglobin
    Biospecimen Aliquots and ComponentsHemolysis
    Fine needle aspiration-induced
    Not induced