Interference by o-phenanthroline in the radioimmunoassay of angiotensin II in small volume blood samples.

Braithwaite JE, Yandle TG, Nicholls MG, Lewis LK

Publication: Clin Biochem (2012) • Vol 45 • Page 168-70PMID: 22040809View on PubMed

Purpose

The purpose of this paper was to determine the effects of enzyme inhibitor cocktail concentration on the measurement of angiotensin II (Ang-II) in blood.

Conclusion

Increasing the concentration of enzyme inhibitor cocktail by decreasing the blood volume from 3 mL to 1 mL led to a 3-fold increase in measured Ang-II levels. The authors found that phenanthroline had a particularly marked effect on Ang-II measurement.

Studies (1)

Purpose

The purpose of this study was to determine the effects of enzyme inhibitor cocktail concentration on the measurement of Ang-II in blood. To determine the effects of inhibitor to blood ratio, 1, 2 or 3 mL aliquots of blood were mixed with 150 uL of enzyme inhibitor.

Summary of Findings

Increasing the concentration of enzyme inhibitor cocktail by decreasing the blood volume from 3 mL to 1 mL led to a 3-fold increase in measured Ang-II levels (p<0.05). The authors found that adding O-phenanthroline to the buffer, but not the other components of the inhibitor cocktail, led to a marked increase in measured Ang-II.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    PeptideRadioimmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationProtease inhibitor
    Cocktail
    Multiple concentrations evaluated