Interference by o-phenanthroline in the radioimmunoassay of angiotensin II in small volume blood samples.
Braithwaite JE, Yandle TG, Nicholls MG, Lewis LK
Purpose
The purpose of this paper was to determine the effects of enzyme inhibitor cocktail concentration on the measurement of angiotensin II (Ang-II) in blood.
Conclusion
Increasing the concentration of enzyme inhibitor cocktail by decreasing the blood volume from 3 mL to 1 mL led to a 3-fold increase in measured Ang-II levels. The authors found that phenanthroline had a particularly marked effect on Ang-II measurement.
Studies (1)
Purpose
The purpose of this study was to determine the effects of enzyme inhibitor cocktail concentration on the measurement of Ang-II in blood. To determine the effects of inhibitor to blood ratio, 1, 2 or 3 mL aliquots of blood were mixed with 150 uL of enzyme inhibitor.
Summary of Findings
Increasing the concentration of enzyme inhibitor cocktail by decreasing the blood volume from 3 mL to 1 mL led to a 3-fold increase in measured Ang-II levels (p<0.05). The authors found that adding O-phenanthroline to the buffer, but not the other components of the inhibitor cocktail, led to a marked increase in measured Ang-II.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
PeptideRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Protease inhibitor CocktailMultiple concentrations evaluated