Skip to main content

The concentration of plasma ethanol measured with an enzymatic assay is decreased in hemolyzed specimens.

Lippi G, Mercadanti M, Musa R, Aloe R, Cervellin G

Publication: Clin Chim Acta (2012) • Vol 413 • Page 356-7PMID: 22008703View on PubMed

Purpose

The purpose of this paper was to determine the effects of mechanically induced hemolysis on ethanol measurements in plasma from emergency room patients.

Conclusion

Plasma ethanol measurements decreased significantly as hemolysis index and free hemoglobin levels increased.

Studies (1)

Purpose

The purpose of this study was to determine the effects of mechanically induced hemolysis on ethanol measurements in plasma from emergency room patients. Specimens were separated into three aliquots, and the first was centrifuged and analyzed while the second and third aliquots were aspirated through a 0.5 mL syringe equipped with a 30 gauge needle 1 and 2 times, respectively, inducing hemolysis prior to centrifugation and analysis.

Summary of Findings

Plasma ethanol measurements decreased significantly as hemolysis index and free hemoglobin levels increased. The median decreases in plasma ethanol between the non-hemolyzed specimens and those aspirated through a thin needle once and twice were 8% and 15%, respectively.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    Cell count/volumeSpectrophotometry
    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    17-22 g/L cell-free hemoglobin
    38-46 g/L cell-free hemoglobin
    <0.5 g/L cell-free hemoglobin
    Biospecimen Aliquots and ComponentsHemolysis
    Fine needle aspiration-induced
    Not induced