The concentration of plasma ethanol measured with an enzymatic assay is decreased in hemolyzed specimens.
Lippi G, Mercadanti M, Musa R, Aloe R, Cervellin G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically induced hemolysis on ethanol measurements in plasma from emergency room patients. Specimens were separated into three aliquots, and the first was centrifuged and analyzed while the second and third aliquots were aspirated through a 0.5 mL syringe equipped with a 30 gauge needle 1 and 2 times, respectively, inducing hemolysis prior to centrifugation and analysis.
Summary of Findings
Plasma ethanol measurements decreased significantly as hemolysis index and free hemoglobin levels increased. The median decreases in plasma ethanol between the non-hemolyzed specimens and those aspirated through a thin needle once and twice were 8% and 15%, respectively.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerCell count/volumeSpectrophotometryProteinSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 17-22 g/L cell-free hemoglobin38-46 g/L cell-free hemoglobin<0.5 g/L cell-free hemoglobinBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot induced