Skip to main content

Influence of primary sample mixing on routine coagulation testing.

Lippi G, Salvagno GL, Montagnana M, Guidi GC

Publication: Blood Coagul Fibrinolysis (2007) • Vol 18 • Page 709-11PMID: 17890963View on PubMed

Purpose

The purpose of this paper was to determine the effects of specimen mixing and aliquot sequential collection on activated partial thromboplastin time (APTT), prothrombin time (PT), D-dimer, and fibrinogen.

Conclusion

For specimens collected from healthy volunteers into Terumo Europe siliconized buffered trisodium citrate tubes (liquid additives), the number of inversions immediately after collection (0, 6, or 12) and aliquot sequential collection did not significantly affect APTT, PT, D-dimer, or fibrinogen levels.

Studies (1)

Purpose

The purpose of this study was to determine the effects of specimen mixing and aliquot sequential collection on APTT, PT, D-dimer, and fibrinogen. Three consecutive specimens were collected from a single venipuncture in Terumo siliconized buffered trisodium citrate tubes. The first aliquot was left in a vertical position, the second aliquot was inverted 6 times, and the third aliquot was inverted 12 times prior to centrifugation and analysis 30-60 min later.

Summary of Findings

For specimens collected from healthy volunteers into Terumo Europe siliconized buffered trisodium citrate tubes (liquid additives), the number of inversions immediately after collection (0, 6, or 12) and aliquot sequential collection did not significantly affect APTT, PT, D-dimer, or fibrinogen levels.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    MorphologyHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
    PeptideHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    3rd collection
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Inverted 12 times
    Inverted 6 times
    Not inverted