HIV-1 drug resistance genotyping from dried blood spots stored for 1 year at 4 degrees C.

Youngpairoj AS, Masciotra S, Garrido C, Zahonero N, de Mendoza C, García-Lerma JG

Publication: J Antimicrob Chemother (2008) • Vol 61 • Page 1217-20PMID: 18344550View on PubMed

Purpose

The purpose of this paper was to determine the effects of storing dried blood spots (DBS) and the amplification method of HIV-1 pol gene from DBS.

Conclusion

Using ViroSeq, only 23 of the 40 DBS stored at 4 degrees C for 12 months were successfully amplified, but all 40 DBS stored at -20 degrees C for 6 months were successfully amplified with Viroseq. Using the in-house amplification method, 38 of the 40 DBS stored at 4 degrees C for 12 months had amplifiable Pol-1. 22 of the 23 DBS amplified successfully with Viroseq after storage at 4 degrees C for 12 months had viral loads in the fresh plasma of more than 10,000 copies/mL. When there was evidence of humidity during storage of DBS for 12 months at 4 degrees C, the ViroSeq assay failed to amplify specimens which had fresh plasma viral loads of less than 14,000 copies/mL.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage temperature and duration of DBS and amplification method of the HIV-1 pol gene. The viral load was determined in fresh plasma prior to storage of DBS. ViroSeq genotyping is dependent on amplification of a 1.8 kb region of the pol gene and the in-house method depended on amplification of a 1 kb region of the pol gene. DBS were stored at -20 degrees C for 6 months (ViroSeq only) and 4 degrees C for 12 months (ViroSeq and In-house method).

Summary of Findings

Only 23 of the 40 DBS stored at 4 degrees C for 12 months were successfully genotyped using ViroSeq, but all 40 DBS stored at -20 degrees C for 6 months were successfully genotyped with Viroseq. Using the in-house method, 38 of the 40 DBS stored at 4 degrees C for 12 months had amplifiable Pol-1. 22 of the 23 DBS successfully amplified with Viroseq after storage at 4 degrees C for 12 months had viral loads in the fresh plasma of more than 10,000 copies/mL. When there was evidence of humidity during storage of DBS for 12 months at 4 degrees C, the ViroSeq assay failed to amplify specimens which had fresh plasma viral loads of less than 14,000 copies/mL. Serial dilution of RNA extracts showed decreased amplification of the larger fragment of pol required for the ViroSeq assay which the authors believe indicates degradation of the RNA during storage at 4 degrees C may be responsible for the failures of the ViroSeq assay.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Other PreservativeNone (Fresh)
  • Diagnoses

    AIDS/HIV-related
  • Platforms

    RNART-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Air-dried
    None (fresh)
    StorageStorage conditions
    Desiccant showed evidence of humidity
    No evidence of humidity
    StorageStorage duration
    0 months
    12 months
    6 months
    StorageStorage temperature
    -20 degrees C
    4 degrees C
    RT-PCR SpecificTargeted nucleic acid
    Pol
    RT-PCR SpecificTechnology platform
    In-house method
    ViroSeq method