Evaluation of Different Mixing Procedures for K2 EDTA Primary Samples on Hematological Testing

Lippi G, Salvagno G L, Montagnana M, Banfi G, Guidi G C

Publication: Labmedicine (2007) • Vol 38 • Page 723-725

Purpose

The purpose of this paper was to determine the effects of specimen mixing and aliquot sequential collection on 7 hematological parameters.

Conclusion

Although some statistically significant changes in hematological parameters were observed between specimens collected 1st and not inverted and those collected 2nd and inverted 6 times, none of the changes exceeded acceptable limits of bias derived from biological variation.

Studies (1)

Purpose

The purpose of this study was to determine the effects of specimen mixing and aliquot sequential collection on red blood cell (RBC), white blood cell (WBC), and platelet counts (PLT), hemoglobin, mean cell hemoglobin (MHC), mean cell volume (MCV), hematocrit, mean hemoglobin content (MHC), and mean platelet volume (MPV). Three consecutive specimens were collected from a single venipuncture in K2EDTA tubes. The first aliquot was left in a vertical position, the second aliquot was inverted 6 times, and the third aliquot was inverted 12 times prior to analysis on the Advia 120 hematology analyzer 30-60 min later. The analyzer was equipped with an autosampler that inverted all primary tubes several times just prior to analysis.

Summary of Findings

While no statistically significant differences were observed between the 2nd and 3rd collections which were initially inverted 6 and 12 times, respectively, statistically significant decreases in RBC, hemoglobin, hematocrit, and PLT, and a statistically significant increase in MPV were observed in the 1st collection (not inverted) compared to the 2nd collection (inverted 6 times). However, none of these changes exceeded acceptable biological variation bias. Platelet clumping was not a factor among any specimens in this study.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    3rd collection
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Inverted 12 times
    Inverted 6 times
    Not inverted