Sources of Variability Between Biobanks in the Estimation of DNA Concentration

Jury F, Morris J, Davis E, Ollier W E, Yuille M

Publication: Biopreservation and Biobanking (2012) • Vol 10 • Page 55-61

Purpose

The purpose of this paper was to determine the effects of several analytical variables on the accuracy of DNA concentration measurement in samples extracted from pooled blood.

Conclusion

With qPCR results taken as the standard, absorption spectroscopy was more accurate than fluorescence spectroscopy for the measurement of DNA concentration, the latter giving lower measurements. In particular, use of a Nanodrop for absorption spectroscopy, as opposed to other instruments, led to increased accuracy in DNA concentration measurement. Importantly, the authors state that centers using the Nanodrop for absorption spectroscopy were less likely to dilute their samples, and sample dilution was associated with decreased accuracy for absorption spectroscopy. The authors state that sample dilution did not significantly affect the accuracy of fluorescence spectroscopy. Averaging two or more measurements was associated with increased accuracy for fluorescence spectroscopy, and this variable was not addressed for absorption spectroscopy. Mixing time of the DNA sample and the time the liquid handling device was last serviced did not affect the accuracy of DNA concentration measurement overall, but the authors state that these variables may possibly influence the two technologies differently.

Studies (1)

Purpose

The purpose of this study was to determine the effects of several analytical variables on the accuracy of DNA concentration measurement in samples extracted by phenol/chloroform from pooled blood. 72 centers participated in measuring DNA concentrations of 8 stock solutions (prepared from the same 45-specimen pool) by their usual methods. The stock solutions consisted of 8, 31, 77, 137, 178, 222, 236, and 335 ng DNA/uL as measured by real time qPCR.

Summary of Findings

With qPCR results taken as the standard, absorption spectroscopy was more accurate than fluorescence spectroscopy for the measurement of DNA concentration, the latter giving lower measurements. In particular, use of a Nanodrop for absorption spectroscopy, as opposed to other instruments, led to increased accuracy in DNA concentration measurement. Importantly, the authors state that centers using the Nanodrop for absorption spectroscopy were less likely to dilute their samples, and sample dilution was associated with decreased accuracy for absorption spectroscopy. The authors state that sample dilution did not significantly affect the accuracy of fluorescence spectroscopy. Averaging two or more measurements was associated with increased accuracy for fluorescence spectroscopy, and this variable was not addressed for absorption spectroscopy. The authors state that mixing time of the DNA sample and the time the liquid handling device was last serviced did not affect the accuracy of DNA concentration measurement overall, but the authors suggest that these variables may possibly influence the two technologies differently. Interestingly, the authors state that all data sets from centers performing fluorescence microscopy indicated that samples were mixed, while 16% of absorption spectroscopy data sets were from unmixed samples. The authors state that there were no significant differences between the use of the Picogreen broad range kit and the Picogreen narrow range kit or between different excitation frequencies, but measurements taken at lower emission frequencies were associated with increased accuracy.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    DNAReal-time qPCR
    DNASpectrophotometry
    DNAFluorometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationSample mixing
    1-5 min
    16-30 min
    31-60 min
    6-15 min
    <1 min
    No mixing
    Overnight
    Variable durations
    Fluorometry SpecificData handling
    One measurement
    Two or more measurements averaged
    Fluorometry SpecificDetection method
    Picogreen broad range kit
    Picogreen narrow range kit
    Unspecified range of emission frequencies
    Unspecified range of excitation frequencies
    Fluorometry SpecificFrequency of servicing
    More than 18 months ago
    Never
    Not applicable
    Unknown
    Within 13-18 months
    Within 6 months
    Within 7-12 months
    Fluorometry SpecificTemplate/input amount
    Diluted sample
    Undiluted sample
    Spectrophotometry SpecificData handling
    One measurement
    Two or more measurements averaged
    Spectrophotometry SpecificFrequency of servicing
    More than 18 months ago
    Never
    Not applicable
    Unknown
    Within 13-18 months
    Within 6 months
    Within 7-12 months
    Spectrophotometry SpecificTechnology platform
    17 instruments compared
    Spectrophotometry SpecificTemplate/input amount
    Diluted sample
    Undiluted sample
    Real-time qPCR SpecificTechnology platform
    Fluoremetry
    Spectrophotometry