No influence of a butterfly device on routine coagulation assays and D-dimer measurement.
Lippi G, Salvagno GL, Guidi GC
Purpose
The purpose of this paper was to determine the effects of using a butterfly device and extension tubing rather than a straight needle for blood collection on routine coagulation testing.
Conclusion
Prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, and D-dimer measurements were not significantly different between specimens collected using a 21 gauge straight needle and those collected using a 21 gauge needle butterfly device with 300 mM grade polyvinyl chloride (PVC) tubing and a Luer adapter.
Studies (1)
Purpose
The purpose of this study was to determine the effects of using a butterfly device and extension tubing rather than a straight needle for blood collection on routine coagulation testing.
Summary of Findings
PT, APTT, fibrinogen, and D-dimer measurements were not significantly different between specimens collected using a 21 gauge straight needle and those collected using a 21 gauge needle butterfly device with 300 mM grade polyvinyl chloride (PVC) tubing and a Luer adapter.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerGlycoproteinHematology/ auto analyzerPeptideELISAPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Butterfly deviceExtension tubingNeedleNo extension tubing