No influence of a butterfly device on routine coagulation assays and D-dimer measurement.

Lippi G, Salvagno GL, Guidi GC

Publication: J Thromb Haemost (2005) • Vol 3 • Page 389-91PMID: 15670053View on PubMed

Purpose

The purpose of this paper was to determine the effects of using a butterfly device and extension tubing rather than a straight needle for blood collection on routine coagulation testing.

Conclusion

Prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, and D-dimer measurements were not significantly different between specimens collected using a 21 gauge straight needle and those collected using a 21 gauge needle butterfly device with 300 mM grade polyvinyl chloride (PVC) tubing and a Luer adapter.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using a butterfly device and extension tubing rather than a straight needle for blood collection on routine coagulation testing.

Summary of Findings

PT, APTT, fibrinogen, and D-dimer measurements were not significantly different between specimens collected using a 21 gauge straight needle and those collected using a 21 gauge needle butterfly device with 300 mM grade polyvinyl chloride (PVC) tubing and a Luer adapter.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    MorphologyHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
    PeptideELISA
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionMethod of fluid acquisition
    Butterfly device
    Extension tubing
    Needle
    No extension tubing