Evaluation of a computer-directed pneumatic-tube system for pneumatic transport of blood specimens.
Weaver DK, Miller D, Leventhal EA, Tropeano V
Purpose
The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport on 15 clinical chemistry and 6 hematological analytes in blood.
Conclusion
Lactate dehydrogenase (LD) levels were 5.9-8% higher after transport by PTS than after hand-delivery and were also increased when the tubes were only 50 or 75% filled compared to 100% full tubes. After PTS transport, levels of all other analytes remained within one standard deviation of the mean measured after hand-delivery and within the assay precision.
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS transport on 15 clinical chemistry and 6 hematological analytes and the effects of tube fill volume on LD levels in blood. Specimens were transported up to 253 m with 16 bends at 7.6 m/s in carriers with foam inserts which prevented movement of the evacuated glass tubes.
Summary of Findings
LD levels were 5.9-8% higher after transport by PTS than after hand-delivery and were also increased when the tubes were only 50 or 75% filled compared to 100% full tubes. After PTS transport, levels of sodium, potassium, chloride, carbon dioxide (CO2), total protein, albumin, calcium, glucose, creatinine, total bilirubin, alkaline phosphatase, aspartate transaminase, acid phosphatase, uric acid, leukocyte count, erythrocyte count, hemoglobin, hematocrit, prothrombin time (PT) and partial thromboplastin time (PTT) remained within one standard deviation of the mean measured after hand-delivery and within the assay precision.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerGasClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinHematology/ auto analyzerMorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 50% full tube75% full tubeFull tubeStorage Within hospital transportation method Hand-deliveredPneumatic tube system