The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport and transport speed on indicators of hemolysis in blood specimens.
Speed of sample transportation by a pneumatic tube system can influence the degree of hemolysis.
Tiwari AK, Pandey P, Dixit S, Raina V
Purpose
Conclusion
PTS transport at high speed (3 m/s) for 115 m resulted in increased lactate dehydrogenase (LD), and high speed transport for 225 m resulted in increased hemoglobin, potassium (K) and LD compared to hand-delivery, indicating high speed PTS transport induced hemolysis. PTS transport at low speed (2 m/s) for 115 m did not affect LD, K or hemoglobin compared to specimens that were hand-delivered.
Studies (1)
Purpose
Summary of Findings
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Specimen transport duration/condition 115 m at 2 m/s115 m at 3 m/s225 m at 3 m/sStorage Within hospital transportation method Hand-deliveredPneumatic tube system