Blood gas analysis: POCT versus central laboratory on samples sent by a pneumatic tube system.

Zaman Z, Demedts M

Publication: Clin Chim Acta (2001) • Vol 307 • Page 101-6PMID: 11369343View on PubMed

Purpose

The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport and blood gas analyzer type on the measurement of blood gas in specimens from patients with lung diseases.

Conclusion

After PTS transport, there were small changes in pH and partial pressure carbon-dioxide (pCO2)compared to specimens that were not transported, and there was a large bias in partial pressure oxygen (pO2) when specimens containing air bubbles were transported by PTS compared to those analyzed without being transported.

Studies (1)

Purpose

The purpose of this study was to determine the effects of PTS transport and blood gas analyzer type on the measurement of blood gas in specimens from patients with lung diseases. Specimens were analyzed at bedside using the IL blood gas analyzer and then sent at room temperature through a PTS and reanalyzed on the Corning 288 blood gas analyzer. This second analysis occurred an average of 13.3 min later (<10 min transport) in specimens from the lung function laboratory (LFL) and 20.2 min (<4 min transport) later in specimens from the pneumology ward.

Summary of Findings

After PTS transport, specimens from the LFL had a 0.0012 increase in pH and a 0.29 mm mercury (Hg) increase in pCO2 compared to specimens analyzed at the patient bedside. The average pO2 in specimens from patients in the LFL undergoing a 100% oxygen (O2) test was 446 mm Hg when analyzed in the LFL and 346 mm Hg after PTS transport to the central laboratory. Prior to the 100% O2 test, LFL patient specimens contained no air bubbles, had a bedside PO2 of 41-107 mm Hg, and showed an average decrease in PO2 of 2.4 mm Hg after PTS transport to the central laboratory. Specimens from the pneumology ward frequently had air bubbles (68%), and after PTS transport, had average increases in pH of 0.012, decreases in pCO2 of 1.64 mm Hg, and increases in PO2 of 12.93 mm Hg (initially 37-183 mm Hg) compared to specimens that were analyzed immediately. Importantly, the effects attributed to PTS transport cannot be distinguished from those due to use of different analyzers or delayed analysis.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Other diagnoses
  • Platforms

    GasClinical chemistry/auto analyzer
    Small moleculepH
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Lung disease
    StorageSpecimen transport duration/condition
    0 min
    <10 min
    <4 min
    Air bubbles
    No air bubbles
    StorageWithin hospital transportation method
    Not transported
    Pneumatic tube system
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Corning 288 blood gas analyzer
    IL blood gas analyzer