A comparison of syringes to collect blood for analysis of gases, electrolytes and glucose.
Carter BG, Tibballs J, Hochmann M, Osborne A, Chiriano A, Murray G
Purpose
The purpose of this paper was to determine the effects of collection syringe type and specimen volume on blood gas, electrolyte, and glucose analysis.
Conclusion
There were no clinically significant differences in pH, sodium, potassium, glucose, ionized calcium, partial pressure oxygen (PO2), or partial pressure carbon dioxide (PCO2) between specimens collected with Johns, Marksman, or Radiometer syringes. However, the authors state that the differences in blood gas analysis between specimens collected in Marksman versus Johns syringes may be clinically significant for specimens with PO2 values less than 100 mmHg. The authors do not provide statistical analysis comparing draw volumes of 0.3-0.5 mL and 1 mL, however, analyte levels, and the differences between syringe types appear similar for each specimen volume.
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection syringe type and specimen volume on PCO2, PO2, pH, sodium, potassium, glucose, and ionized calcium in blood specimens from children in intensive care. Visible air bubbles were immediately expelled from specimens, and those with remaining visible air bubble contamination were excluded at the time of analysis. Specimens were collected in random order, but all those collected in Marksman syringes were analyzed after those collected in Johns syringes. Batch analysis was performed introducing variations in the time specimens spent at room temperature until analysis.
Summary of Findings
There were no clinically significant differences in pH, sodium, potassium, glucose, or ionized calcium between specimens collected with Johns, Marksman, or Radiometer syringes. Specimens collected in Marksman syringes tended to have higher PO2 and PCO2 values than those collected in Johns syringes, and the differences between the syringe types tended to increase with increasing mean PO2 and PCO2 values. Although these differences were not clinically significant, the authors state that the differences in blood gas analysis between specimens collected in Marksman versus Johns syringes may be clinically significant for specimens with PO2 values less than 100 mmHg. The authors do not provide statistical analysis comparing draw volumes of 0.3-0.5 mL and 1 mL, however, analyte levels, and the differences between syringe types appear similar for each specimen volume.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Johns syringeMarksman syringeRadiometer syringeBiospecimen Aliquots and Components Aliquot size/volume 0.3-0.5 mL1 mL