Serum bicarbonate declines with sample size in vacutainer tubes.
Herr RD, Swanson T
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of tube fill volume, vacutainer size, aspiration into a syringe versus directly into a vacutainer, and delaying analysis by up to 40 minutes on serum bicarbonate levels.
Summary of Findings
There were no differences in bicarbonate levels between specimens aspirated into a syringe and injected into a vacutainer and those aspirated directly into a vacutainer. Further, regardless of vacutainer size, delays of up to 40 minutes before analysis had no effects on serum bicarbonate levels when compared to specimens analyzed immediately after clot formation (10 min after collection). Decreasing fill volumes significantly decreased serum bicarbonate levels for both sizes of vacutainer tubes, regardless of any delays before analysis.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Small moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Syringe drawVacuum tubeBiospecimen Acquisition Type of collection container/solution 10 mL vacutainer3 mL vacutainerBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1 mL3 mL6 mL9.5 mLStorage Time at room temperature 10 min20 min30 min40 min