Implementation of a pneumatic-tube system for transport of blood specimens.
Poznanski W, Smith F, Bodley F
Purpose
The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport, tube fill volume, anticoagulation with heparin, and clotting prior to transport on clinical chemistry analytes in blood.
Conclusion
For specimens collected with no anticoagulant, total carbon dioxide (CO2) and chloride (Cl-) were slightly increased in specimens transported by PTS rather than hand-delivered. PTS transport rather than hand-delivery did not affect potassium (K+), hemoglobin, or lactate dehydrogenase (LDH) when the collection tubes were half-filled or contained heparin, but LDH was increased in clotted blood that was sent by PTS rather than hand-delivered.
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS transport, tube fill volume, anticoagulation with heparin, and clotting prior to transport on clinical chemistry analytes in blood. Blood was obtained from patients in the emergency room or intensive care unit and transported immediately or allowed to clot for 15 min prior to transport. For blood gas analysis the specimens were wrapped in a cold pack prior to immediate transport.
Summary of Findings
CO2 and Cl- were slightly increased in specimens transported by PTS rather than hand-delivered. Transport method did not affect K+, hemoglobin, or LDH when collection tubes were half-filled or when they contained heparin, but LDH was increased in clotted blood that was sent by PTS rather than hand-delivered. For unclotted specimens collected with no anticoagulant, glucose, urea, sodium, K+, LDH, hemoglobin, pH, partial pressure oxygen (PO2), and partial pressure carbon dioxide (PCO2) were unaffected by transport method.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant HeparinNoneBiospecimen Aliquots and Components Aliquot size/volume Filled vacutainerHalf-filled vacutainerStorage Specimen transport duration/condition ClottedUnclottedStorage Within hospital transportation method Hand-deliveredPneumatic tube system