Skip to main content

Preanalytical effects of pneumatic tube transport on routine haematology, coagulation parameters, platelet function and global coagulation.

Wallin O, Söderberg J, Grankvist K, Jonsson PA, Hultdin J

Publication: Clin Chem Lab Med (2008) • Vol 46 • Page 1443-9PMID: 18844500View on PubMed

Purpose

The purpose of this paper was to determine the effects of aspirin therapy and pneumatic tube system (PTS) transport on coagulation parameters and cell counts in venous blood.

Conclusion

Small differences in reticulocytes, high fluorescent reticulocytes, platelet counts, platelet distribution width, and time to clot formation were observed when specimens from subjects not taking aspirin were transported by PTS rather than analyzed immediately. In specimens collected after patients had taken aspirin for 1 week, reticulocytes, high fluorescent reticulocytes, reticulocyte hemoglobin equivalent, platelet distribution width, antithrombin percent, coagulation index, and time to clot formation were significantly different in PTS-transported specimens compared to specimens that were not transported.

Studies (1)

Purpose

The purpose of this study was to determine the effects of aspirin therapy and PTS transport on 6 coagulation and 19 cell count parameters in venous blood from healthy subjects. Specimens transported by PTS made a round trip to and from the transfer system 500 m away (155 sec) prior to analysis. Blood was anticoagulated with EDTA or citrate.

Summary of Findings

Small differences in reticulocytes (-3.6%, p=0.012), high fluorescent reticulocytes (-50%, p=0.003), platelet counts (1.2%, p=0.006), platelet distribution width (-2.5%, p=0.008), and time to clot formation (-16%, p=0.037) were observed when specimens from subjects not taking aspirin were transported by PTS rather than analyzed immediately. In specimens collected after patients had taken aspirin for 1 week, reticulocytes (-8.1%, p=0.009), high fluorescent reticulocytes (-52%, p=0.034), reticulocyte hemoglobin equivalent (-1.4%, p=0.001), platelet distribution width (-0.8%, p=0.047), antithrombin percent (0.7%, p=0.026), coagulation index (60%, p=0.043), and time to clot formation (-22%, p=0.043) were significantly different in PTS-transported specimens compared to specimens that were not transported. The closure time (CT) using the collagen membrane coated with epinephrine (CEPI) cartridge showed a large difference between specimens from subjects not taking aspirin transported by PTS rather than not transported , but the differences were driven by a few individuals with large increases or decreases. The CT using the adenine diphosphate (ADP) cartridge (CADP) was not different in specimens transported by PTS compared to specimens not transported.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    GlycoproteinHematology/ auto analyzer
    MorphologyHematology/ auto analyzer
    PeptideHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionOther drugs
    After 1 week of aspirin therapy
    Prior to aspirin therapy
    StorageWithin hospital transportation method
    Not transported
    Pneumatic tube system