Pneumatic tube system induced haemolysis: assessing sample type susceptibility to haemolysis.
Sodi R, Darn SM, Stott A
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection container and PTS transport on hemolysis in blood specimens.
Summary of Findings
The median HI of serum collected in plain serum tubes and transported by PTS was 37, which exceeded the hemolysis threshold of 31 and was significantly higher than the median HI of 4 observed in serum collected in plain serum tubes and transported by porter (p<0.0001). The HI of specimens transported by PTS in 13 mm serum-gel (10), and K-EDTA(6) were significantly higher than when transported by courier (6, p=0.0367, and 2, p=0.0001, respectively). The HI of specimens collected in fluoride-EDTA tubes and transported by PTS were significantly lower than when hand-delivered (6 versus 13, p=0.0021). Significantly higher HI were observed when specimens were collected in 13 mm serum gel tubes rather than 16 mm serum gel tubes and transported by PTS (p=0.046) or hand-delivered (p=0.013).
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Fluoride-EDTALithium heparinNonePotassium EDTABiospecimen Acquisition Type of collection container/solution 13 mm serum-gel tube16 mm serum-gel tubePlain serum tubeStorage Within hospital transportation method Hand-deliveredPneumatic tube system