Sample evaporation and its impact on the operating performance of an automated selective-access analytical system.
Burtis CA
Purpose
The purpose of this paper was to determine the effects of sample cup size, specimen volume, capping sample cups, and storage time on the measurement of glucose in serum by the Dimension auto analyzer.
Conclusion
Glucose concentration increased with storage duration when serum was held in sample cups prior to analysis in the Dimension auto analyzer, and these increases were greatest when the volume was small. Capping the sample cups attenuated these increases, especially for the serum held in larger sample cups.
Studies (1)
Purpose
The purpose of this study was to determine the effects of sample cup size, specimen volume, capping sample cups, and storage time on the measurement of glucose in serum by the Dimension auto analyzer. Pooled serum from an unspecified origin was used in these studies.
Summary of Findings
Glucose concentration increased with storage duration (6.3-16.5 mg/L increases after 4 h) when serum was held in sample cups prior to analysis in the Dimension auto analyzer. Glucose concentration increased more in aliquots with a smaller volume. Capping the sample cups attenuated these increases in glucose concentration (2.3-13.1 mg/L increases after 4 h), especially for serum held in 2.0 mL cups rather than in 0.5 mL.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 0.125 mL0.250 mL0.375 mL0.5 mL1.0 mL1.5 mL2.0 mLStorage Storage conditions CappedNot cappedStorage Time at room temperature 0 h10 h2 h4 h6 h8 hStorage Type of storage container 0.5 mL sample cup2.0 mL sample cup