Pseudometabolic acidosis caused by underfill of Vacutainer tubes.
Herr RD, Swanson T
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of specimen volume and venting Vacutainer tubes after specimen collection on serum bicarbonate and anion gap values in specimens from emergency department patients. The authors state that time to analysis was not controlled but reflected actual patient care.
Summary of Findings
In both vented and unvented tubes, mean serum bicarbonate values decreased and mean anion gap increased significantly as specimen volume decreased from 10 mL to 1 mL. No significant differences in serum bicarbonate or anion gap were found between vented and unvented tubes.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 1 mL10 mL3 mLStorage Storage conditions Unvented VacutainerVacutainer vented after collection