Effects of a pneumatic tube system on routine and novel hematology and coagulation parameters in healthy volunteers.
Kratz A, Salem RO, Van Cott EM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS transport on hematological and coagulation parameters in blood. Cell counts were obtained from potassium-EDTA anticoagulated blood within 1 h of venipuncture. Citrated blood was centrifuged and plasma was analyzed for coagulation parameters within 2 h or was frozen at -70 degrees C until analysis for coagulation assays.
Summary of Findings
Transport of specimens by PTS rather than by hand-delivery did not affect CBCs, white blood cell differentials, PT, APTT, levels of fibrinogen or fibrin monomers, or ESR. The average MPC was slightly higher (28.1 versus 28.2 g/dL) in PTS-transported specimens than in hand-delivered specimens, but this difference was only significant when the two specimens from individuals taking aspirin were excluded (p=0.04). While 18 specimens had higher MPC after PTS transport than after hand-delivery, 10 specimens showed lower MPC after PTS transport, and 5 had the same MPC after PTS transport as after hand-delivery.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)FrozenDiagnoses
NormalPlatforms
Cell count/volumeHematology/ auto analyzerGlycoproteinHematology/ auto analyzerMorphologyHematology/ auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation FrozenNone (fresh)Storage Within hospital transportation method Hand-deliveredPneumatic tube system