Evaluation of the relationship between capillary and venous plasma glucose concentrations obtained by the HemoCue Glucose 201+ system during an oral glucose tolerance test.
Ignell C, Berntorp K
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of using capillary versus venous blood specimens for the measurement of glucose levels in women undergoing an oral glucose tolerance test. Capillary or venous whole blood was analyzed in a HemoCue Glucose 201+ Analyzer, and blood glucose concentrations were converted to plasma glucose concentrations using an adjustment factor.
Summary of Findings
Capillary plasma glucose concentrations were significantly higher than venous plasma concentrations at all time points (fasting, 30 min after oral glucose, and 2 h after oral glucose). The biggest average difference (1.8 mmol/L) and lowest correlation (r=0.81) between capillary and venous plasma glucose were found at the 30 min post load time point. The difference between capillary and venous specimens increased with increasing glucose concentrations.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesPlatforms
CarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Previous gestational diabetesBiospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Time of biospecimen collection 2 h after oral glucose consumption30 min after oral glucose consumptionAfter overnight fast