Skip to main content

Influence of blood collection in plastic vs. glass evacuated serum-separator tubes on hormone and tumour marker levels.

Smets EM, Dijkstra-Lagemaat JE, Blankenstein MA

Publication: Clin Chem Lab Med (2004) • Vol 42 • Page 435-9PMID: 15147155View on PubMed

Purpose

The purpose of this paper was to determine the effects of collecting blood in plastic versus glass serum separator tubes (SST) on the measurement of 35 clinical chemistry analytes in serum.

Conclusion

While statistically significant differences were found in free triiodothyronine, progesterone, prolactin, prostate-specific antigen (PSA), and pregnancy-associated plasma protein-A (PAPP-A) between specimens collected in plastic and glass SST, none of the differences were clinically relevant. Further, no significant differences were found in 30 other analytes between specimens collected in the two tube types.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collecting blood in plastic versus glass SST on the measurement of 35 clinical chemistry analytes in serum. While most specimens were collected in random order in both tubes types, some extra specimens were collected in glass SST and then 1 mL aliquots of serum were transferred to plastic SST, kept on a roller bank for 1 hour at room temperature, and analyzed as plastic SST specimens.

Summary of Findings

While statistically significant differences were found in free triiodothyronine, progesterone, prolactin, PSA, and PAPP-A between specimens collected in plastic and glass SST, none of the differences were clinically relevant. Further, no significant differences were found in 30 other analytes between specimens collected in the two tube types.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    GlycoproteinClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass SST
    Plastic SST II Plus