Pneumatic transport exacerbates interference of room air contamination in blood gas samples.
Astles JR, Lubarsky D, Loun B, Sedor FA, Toffaletti JG
Purpose
The purpose of this paper was to determine the effects of pneumatic transfer system (PTS) transport on blood partial pressure oxygen (PO2).
Conclusion
Air bubbles did not affect the PO2 when specimens were left undisturbed for 5 min, but they significantly increased the PO2 when specimens with less than 100 mm mercury (Hg) oxygen (O2) were transported by PTS and decreased the PO2 when specimens with 339 mm Hg O2 were transported by PTS. The changes could be partially attenuated either by slowing the speed of transport by 50% or, for specimens with an initial PO2 of 342 mm Hg O2, by transporting them in a bag filled with an ice slurry, but not water.
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS transport conditions on the PO2 of tonometered blood and blood from normal and hypoxemic patients.
Summary of Findings
Air bubbles did not affect the PO2 when specimens were left undisturbed for 5 min or when specimens tonometered with 142 mm Hg O2 were transported by PTS, but they significantly increased the PO2 when specimens with less than 100 mm Hg O2 were transported by PTS and decreased the PO2 when specimens with 339 mm Hg O2 were transported by PTS. Importantly, air bubbles had a significant effect on the PO2 of specimens from both normal and hypoxemic patients. The addition of a liner during PTS did not provide protection against the changes in PO2. Removal of air from the liner bag prior to transport significantly reduced the PO2. Filling the liner bag with an ice slurry stabilized the PO2 of specimens with 342 mm Hg O2 during PTS transport but did not prevent changes in specimens with a PO2 of 81 mm Hg O2. Reducing the speed of PTS to 50% significantly reduced the effects of PTS transport on the PO2.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedOther diagnosesPlatforms
GasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Post-cardiac surgery intensive care patientsUnspecified elective surgical patientUnspecified pneumatic diseaseBiospecimen Aliquots and Components Biospecimen components 0.2 mL air bubble0.5 mL air bubble142 mm Hg O2339 mm Hg O265 mm Hg O275 mm Hg O2Storage Specimen transport duration/condition Pneumatic transfer in ice slurry filled bagPneumatic transfer in water filled bagPneumatic transport at 50% of normal speedPneumatic transport at 75% of normal speedPneumatic transport with liner bagPneumatic transport without liner bagStorage Within hospital transportation method Not transportedPneumatic tube system