Skip to main content

Pneumatic transport exacerbates interference of room air contamination in blood gas samples.

Astles JR, Lubarsky D, Loun B, Sedor FA, Toffaletti JG

Publication: Arch Pathol Lab Med (1996) • Vol 120 • Page 642-7PMID: 8757468View on PubMed

Purpose

The purpose of this paper was to determine the effects of pneumatic transfer system (PTS) transport on blood partial pressure oxygen (PO2).

Conclusion

Air bubbles did not affect the PO2 when specimens were left undisturbed for 5 min, but they significantly increased the PO2 when specimens with less than 100 mm mercury (Hg) oxygen (O2) were transported by PTS and decreased the PO2 when specimens with 339 mm Hg O2 were transported by PTS. The changes could be partially attenuated either by slowing the speed of transport by 50% or, for specimens with an initial PO2 of 342 mm Hg O2, by transporting them in a bag filled with an ice slurry, but not water.

Studies (1)

Purpose

The purpose of this study was to determine the effects of PTS transport conditions on the PO2 of tonometered blood and blood from normal and hypoxemic patients.

Summary of Findings

Air bubbles did not affect the PO2 when specimens were left undisturbed for 5 min or when specimens tonometered with 142 mm Hg O2 were transported by PTS, but they significantly increased the PO2 when specimens with less than 100 mm Hg O2 were transported by PTS and decreased the PO2 when specimens with 339 mm Hg O2 were transported by PTS. Importantly, air bubbles had a significant effect on the PO2 of specimens from both normal and hypoxemic patients. The addition of a liner during PTS did not provide protection against the changes in PO2. Removal of air from the liner bag prior to transport significantly reduced the PO2. Filling the liner bag with an ice slurry stabilized the PO2 of specimens with 342 mm Hg O2 during PTS transport but did not prevent changes in specimens with a PO2 of 81 mm Hg O2. Reducing the speed of PTS to 50% significantly reduced the effects of PTS transport on the PO2.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Other diagnoses
  • Platforms

    GasClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Post-cardiac surgery intensive care patients
    Unspecified elective surgical patient
    Unspecified pneumatic disease
    Biospecimen Aliquots and ComponentsBiospecimen components
    0.2 mL air bubble
    0.5 mL air bubble
    142 mm Hg O2
    339 mm Hg O2
    65 mm Hg O2
    75 mm Hg O2
    StorageSpecimen transport duration/condition
    Pneumatic transfer in ice slurry filled bag
    Pneumatic transfer in water filled bag
    Pneumatic transport at 50% of normal speed
    Pneumatic transport at 75% of normal speed
    Pneumatic transport with liner bag
    Pneumatic transport without liner bag
    StorageWithin hospital transportation method
    Not transported
    Pneumatic tube system