Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality.

Fernandes CM, Worster A, Eva K, Hill S, McCallum C

Publication: J Emerg Nurs (2006) • Vol 32 • Page 139-43PMID: 16580476View on PubMed

Purpose

The purpose of this paper was to determine the effects of using a pneumatic tube system (PTS) rather than a courier for specimen transport on hemoglobin and potassium levels in blood specimens.

Conclusion

Using a PTS rather than transfer by human courier had no effects on hemolysis, hemoglobin or potassium levels in blood specimens, but decreases in turnaround time were observed.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using a PTS rather than a courier for specimen transport on hemoglobin and potassium levels in blood specimens. Two different hospitals were included in this study, one of which used a PTS system while the other one used human couriers.

Summary of Findings

Transport of specimens by PTS decreased the average delay to analysis for hemoglobin by 10 min and potassium by 8 min. Using PTS rather than human courier for specimen transport had no effects on hemolysis rates or hemoglobin or potassium levels.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinHematology/ auto analyzer
    MorphologyMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system