Electrolytes assessed by point-of-care testing - Are the values comparable with results obtained from the central laboratory?

Chacko B, Peter JV, Patole S, Fleming JJ, Selvakumar R

Publication: Indian J Crit Care Med (2011) • Vol 15 • Page 24-9PMID: 21633542View on PubMed

Purpose

The purpose of this paper was to determine the effects of analyzing whole blood at the bedside rather than serum after sending whole blood through a pneumatic transfer system (PTS) on sodium and potassium levels.

Conclusion

Clinically relevant decreases in sodium or potassium (when <3 mmol/L) were observed when whole blood was analyzed at the bedside rather than when serum was analyzed after PTS transport.

Studies (1)

Purpose

The purpose of this study was to determine the effects of analyzing whole blood at the bedside rather than analyzing serum after sending whole blood through a PTS on sodium and potassium levels. Specimens for analysis at the bedside were collected into an ABG syringe coated with lithium heparin, while specimens for analysis in the central lab were collected in BD vacutainers for serum.

Summary of Findings

After PTS transport, serum sodium levels, measured in the central lab using the AU2700, were an average of 4 mmol/L higher than whole blood levels measured using a GEM 3000 at the patient bedside (p<0.001). Generally, potassium levels in PTS transported serum and untransported whole blood showed good agreement (Lin concordance correlation, Pc0.96), but when potassium levels were below 3 mmol/L, the serum levels were, on average, 0.6 mmol/L higher in transported specimens, and the agreement between the two was low (Pc=0.53). The differences in levels of potassium (below 3 mmol/L only) or sodium between whole blood analyzed at the bedside and serum analyzed after PTS transport were clinically relevant.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Serum
    Whole blood
    StorageWithin hospital transportation method
    Not transported
    Pneumatic tube system
    Clinical chemistry/auto analyzer SpecificTechnology platform
    AU2700 (indirect ion selective electrode)
    GEM 3000 (direct ion selective electrode)